The purpose of this clinical study is to validate the SpO2 accuracy of the BiOSENCY BORA BAND™ wristband Pulse Oximeter during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. The end goal is to provide supporting documentation for the SpO2 accuracy validation of BORA BAND™ wristband Pulse Oximeter.
It is required that the Accuracy Root Mean Square (ARMS) performance of the BORA BAND™ wristband Pulse Oximeter will meet a specification of 3.5 or better in non-motion conditions for the range of 70 - 100% SpO2 thereby demonstrating an acceptable SpO2 accuracy performance specification.
Condition or disease | Intervention/treatment | Phase |
---|---|---|
Hypoxia | Device: BORA Band SpO2 Validation Study | Not Applicable |
The SpO2 accuracy performance of BiOSENCY BORA BAND™ wristband Pulse Oximeter will be evaluated during non-motion conditions over the range of 70-100% SaO2 and compared to arterial blood samples assessed by CO-Oximetry. A minimum of 10 healthy adult male and female subjects, ranging in age and pigmentation from light to dark, will be enrolled in the study to meet the study design requirements defined by ISO 80601-2-61:2017 and by the FDA's Guidance for Pulse Oximeters (March 4, 2013). The subjects will have an arterial catheter placed in the radial artery to allow for simultaneous blood samples during stable plateaus of induced hypoxic levels. The investigational device will be placed on the wrist opposite for the test per the instructions for use. Simultaneous data collection will be set up for the system under test and control oximeter.
For the data analysis, the control oximeter will be used to assess the stability of each data point. Data that is found to be unstable will be removed prior to the comparative analysis. The CO-Oximeter data will be reviewed to make sure it does not contain any anomalous values such as elevated, low or inconsistent COHb, MetHb, or tHb data. Anomalous values will be removed from the analysis prior to pairing of the SpO2 and SaO2 data. The statistical analysis is performed on a minimum of 200 data points collected from at least 10 subjects for the range of 70% to 100% SaO2 with allowance of ± 3% of the target range. Functional SaO2 as measured by Reference CO-Oximetry will be used as the basis for comparison. The Accuracy Root Mean Square (ARMS) calculation is used to determine the SpO2 accuracy performance. Success will be achieved with an ARMS of 3.5 or better showing equivalence to the Gold Standard Reference CO-Oximetry providing documentation to support SpO2 accuracy claims for the investigational device.
The study population will include 10-15 healthy non-smoking (or has refrained from smoking for 2 days) competent adults, 18-50 years of age. The subject selection will be an equitable distribution of males and females of any race with varying skin tones including at least 2 darkly pigmented subjects or 15% of the subject pool, whichever is larger. The subjects must understand the study and consent to participate by signing the Informed Consent Form. The subjects must be healthy showing no evidence of medical problems as indicated by satisfactorily completing the health assessment form.
Study Type : | Interventional (Clinical Trial) |
Actual Enrollment : | 10 participants |
Allocation: | N/A |
Intervention Model: | Single Group Assignment |
Masking: | None (Open Label) |
Primary Purpose: | Device Feasibility |
Official Title: | SpO2 and Pulse Rate Accuracy Comparison of BiOSENCY BORA Band to Arterial Blood CO-Oximetry and Reference ECG |
Actual Study Start Date : | February 19, 2019 |
Actual Primary Completion Date : | February 21, 2019 |
Actual Study Completion Date : | February 21, 2019 |
Accuracy of the BORA Band SpO2 measurement as compared to reference co-oximetry performed on arterial blood samples.
Results shall include
Accuracy of the BORA Band pulse rate measurement as compared to reference electrocardiogram readings.
Results shall include
Tracking Information | |||||||||
---|---|---|---|---|---|---|---|---|---|
First Submitted Date ICMJE | April 4, 2019 | ||||||||
First Posted Date ICMJE | April 17, 2019 | ||||||||
Last Update Posted Date | April 17, 2019 | ||||||||
Actual Study Start Date ICMJE | February 19, 2019 | ||||||||
Actual Primary Completion Date | February 21, 2019 (Final data collection date for primary outcome measure) | ||||||||
Current Primary Outcome Measures ICMJE |
|
||||||||
Original Primary Outcome Measures ICMJE | Same as current | ||||||||
Change History | No Changes Posted | ||||||||
Current Secondary Outcome Measures ICMJE | Not Provided | ||||||||
Original Secondary Outcome Measures ICMJE | Not Provided | ||||||||
Current Other Pre-specified Outcome Measures | Not Provided | ||||||||
Original Other Pre-specified Outcome Measures | Not Provided | ||||||||
Descriptive Information | |||||||||
Brief Title ICMJE | BiOSENCY BORA Band SpO2 Validation Study | ||||||||
Official Title ICMJE | SpO2 and Pulse Rate Accuracy Comparison of BiOSENCY BORA Band to Arterial Blood CO-Oximetry and Reference ECG | ||||||||
Brief Summary |
The purpose of this clinical study is to validate the SpO2 accuracy of the BiOSENCY BORA BAND™ wristband Pulse Oximeter during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. The end goal is to provide supporting documentation for the SpO2 accuracy validation of BORA BAND™ wristband Pulse Oximeter. It is required that the Accuracy Root Mean Square (ARMS) performance of the BORA BAND™ wristband Pulse Oximeter will meet a specification of 3.5 or better in non-motion conditions for the range of 70 - 100% SpO2 thereby demonstrating an acceptable SpO2 accuracy performance specification. |
||||||||
Detailed Description |
The SpO2 accuracy performance of BiOSENCY BORA BAND™ wristband Pulse Oximeter will be evaluated during non-motion conditions over the range of 70-100% SaO2 and compared to arterial blood samples assessed by CO-Oximetry. A minimum of 10 healthy adult male and female subjects, ranging in age and pigmentation from light to dark, will be enrolled in the study to meet the study design requirements defined by ISO 80601-2-61:2017 and by the FDA's Guidance for Pulse Oximeters (March 4, 2013). The subjects will have an arterial catheter placed in the radial artery to allow for simultaneous blood samples during stable plateaus of induced hypoxic levels. The investigational device will be placed on the wrist opposite for the test per the instructions for use. Simultaneous data collection will be set up for the system under test and control oximeter. For the data analysis, the control oximeter will be used to assess the stability of each data point. Data that is found to be unstable will be removed prior to the comparative analysis. The CO-Oximeter data will be reviewed to make sure it does not contain any anomalous values such as elevated, low or inconsistent COHb, MetHb, or tHb data. Anomalous values will be removed from the analysis prior to pairing of the SpO2 and SaO2 data. The statistical analysis is performed on a minimum of 200 data points collected from at least 10 subjects for the range of 70% to 100% SaO2 with allowance of ± 3% of the target range. Functional SaO2 as measured by Reference CO-Oximetry will be used as the basis for comparison. The Accuracy Root Mean Square (ARMS) calculation is used to determine the SpO2 accuracy performance. Success will be achieved with an ARMS of 3.5 or better showing equivalence to the Gold Standard Reference CO-Oximetry providing documentation to support SpO2 accuracy claims for the investigational device. The study population will include 10-15 healthy non-smoking (or has refrained from smoking for 2 days) competent adults, 18-50 years of age. The subject selection will be an equitable distribution of males and females of any race with varying skin tones including at least 2 darkly pigmented subjects or 15% of the subject pool, whichever is larger. The subjects must understand the study and consent to participate by signing the Informed Consent Form. The subjects must be healthy showing no evidence of medical problems as indicated by satisfactorily completing the health assessment form. |
||||||||
Study Type ICMJE | Interventional | ||||||||
Study Phase ICMJE | Not Applicable | ||||||||
Study Design ICMJE | Allocation: N/A Intervention Model: Single Group Assignment Masking: None (Open Label) Primary Purpose: Device Feasibility |
||||||||
Condition ICMJE | Hypoxia | ||||||||
Intervention ICMJE | Device: BORA Band SpO2 Validation Study
Qualified healthy volunteers will be enrolled into the study. Subjects will recline for the study. Reference sensor(s) will be placed on each subject to evaluate the SpO2 accuracy and performance. Shield material may be used between any adjacent finger sensors to prevent optical crosstalk. Simultaneous data collection will be set up for device under test and the control pulse oximeter. The data from the test device will be collected by the sponsor or trained Clinimark Study staff. Data will be collected for 1 second intervals data analysis. The SpO2 accuracy of the test devices will be evaluated over the oxygen saturation range between 70-100%. The investigational device will only be placed on the wrist. The control pulse oximeter may be alternated through fingers between subjects. |
||||||||
Study Arms ICMJE | Not Provided | ||||||||
Publications * | Not Provided | ||||||||
* Includes publications given by the data provider as well as publications identified by ClinicalTrials.gov Identifier (NCT Number) in Medline. |
|||||||||
Recruitment Information | |||||||||
Recruitment Status ICMJE | Completed | ||||||||
Actual Enrollment ICMJE |
10 | ||||||||
Original Actual Enrollment ICMJE | Same as current | ||||||||
Actual Study Completion Date ICMJE | February 21, 2019 | ||||||||
Actual Primary Completion Date | February 21, 2019 (Final data collection date for primary outcome measure) | ||||||||
Eligibility Criteria ICMJE |
Inclusion Criteria:
Exclusion Criteria:
|
||||||||
Sex/Gender ICMJE |
|
||||||||
Ages ICMJE | 18 Years to 50 Years (Adult) | ||||||||
Accepts Healthy Volunteers ICMJE | Yes | ||||||||
Contacts ICMJE | Contact information is only displayed when the study is recruiting subjects | ||||||||
Listed Location Countries ICMJE | United States | ||||||||
Removed Location Countries | |||||||||
Administrative Information | |||||||||
NCT Number ICMJE | NCT03918018 | ||||||||
Other Study ID Numbers ICMJE | PR2018-312 | ||||||||
Has Data Monitoring Committee | No | ||||||||
U.S. FDA-regulated Product |
|
||||||||
IPD Sharing Statement ICMJE |
|
||||||||
Responsible Party | Biosency | ||||||||
Study Sponsor ICMJE | Biosency | ||||||||
Collaborators ICMJE | Clinimark, LLC | ||||||||
Investigators ICMJE | Not Provided | ||||||||
PRS Account | Biosency | ||||||||
Verification Date | April 2019 | ||||||||
ICMJE Data element required by the International Committee of Medical Journal Editors and the World Health Organization ICTRP |